Dry
Pump Servicing
Specialists
in the supply of Reconditioned Vacuum Equipment

Island
Scientific Ltd offer servicing and repair on DryStar Vacuum Pumps.
We have available full decontamination facilities and all work is
carried out to ISO9001 Quality Standards.
Drystar
Servicing
|
Standard
Service:
Strip
pump apart, decontaminate and clean all components. All contaminants
disposed of according to local environment health requirements
Replace:
PTFE Coating
on claws and sleeves
Bearings
Seals
'O' Rings
Coolant
PFPE fluid in gearbox
Check
and Report if attention needed:
Rotor
and Stator clearances
Electric Motor
Gears
Motor/Pump Coupling
Overhaul
and Replace if necessary:
Thermal
Snap switch
Water Thermal control Valve
Exhaust Manifold
Nitrogen Lines
Final
Assembly:
Leak Test
using Helium
Mass Spectrometer and make leak free
Run and test for 48 hours
Issue final test report
A
Copy of test results included with each service.
|
 |
|
|
 |
Full
servicing of Roots Boosters is also available, (prices on request).
Collection & delivery can be arranged if required, (prices on request).
Contract
servicing rates available on request
| Price
List - Pump Servicing |
Island
Scientific can fully strip and overhaul most makes and types
of Vacuum Pump. From small oil sealed Rotary Pumps to large
Roots Booster Combinations. This includes decontamination of
Pumps from semiconductor applications.
Our Pump service price guide is based on the following work
being undertaken.
|
| Basic
Service for Direct Drive Rotary Pumps. |
- Fully
stripped, cleaned and inspected for wear.
- If
major repairs are needed a quotation will be sent for approval
before the service is completed.
- The
pump motor is then stripped, cleaned and fitted with new
bearings.
- Pump
and motor re-assembled with new 'O' rings, oil seals, gaskets
and oil pump blade.
- Pump
run and fully tested before re-packing and despatch back
to customer.
|
| Major
Service |
| Items charged
on a full overhaul include |
- Bearings
or Bushes (as fitted) in the pump.
- Blades,
springs and pins.
|
We can
also undertake shaft re-building and re-boring of stators at
additional cost.
If the pump has been used on a semiconductor application the
cost of decontamination starts from £75.00/pump depending on
the model.
If down time is your major concern we may be able to offer a
service exchange unit from our extensive range of reconditioned
units.
|
|
Above prices exclude delivery, collection
and VAT.
Prices do not include decontamination
please add for decontamination £90 (approx. dependant on Pump
model).
Pump only decontamination with no
re-build from £150.00.
We will fax you for any extra work
that may be required such as new rotors and stators. If we
find the pump is beyond repair we may be able to supply a
service exchange unit from our existing stocks of reconditioned
equipment.
|
| |
A
Selection of Reconditioned Dry Pumps and Boosters are usually available
from stock.
Tel:
01983 855822
Fax: 01983 852146
E-mail: sales@island-scientific.co.uk
Drystar
is a trademark of B.O.C. Edwards High Vacuum
|